Vertical Zero Datum Program for
NIKON Instruments
NIKON Instruments
- Nikon DTM-A series A5/A10/A20
- Press the (ANG) key while powering on the unit.
- {Tilt Telescope} will appear (index telescope)
- VA: 359.0?} will appear on the display (precisely sight target in face 1) 90 deg. side
- Press the (1) key to execute angle measurement.
- {V2: 180.?}will appear on the display (precisely sight target in face 2) 270deg.side
- Press the (1) key to execute angle measurement.
- Adjustment is complete.
- Nikon DTM-400 series 410/420/430/450
- Nikon DTM-300 series 300/310
- Power the unit on and index telescope. Normal operation mode.
- Press the (FNC) key, then press the (REC) key to enter adjustment mode.
- {Vr: 00.?} will appear on the display (precisely sight target in face 1) 90 deg. side
- Press (ENT) key to execute measurement.
- {Vl: 179.?}will appear on the display (precisely sight target in face 2) 270deg.side
- Press (ENT) key to execute measurement.
- Adjustment is complete. This procedure also adjust the run center of the tilt sensors
- Nikon C100 and D50
- Press the (TRK) and (HOLD) key while powering up the unit.
- {Tilt Telescope} will appear (index telescope)
{AC: 0.0---?? } current correction value. - {V1: 90.?} will appear on the display (precisely sight target in face 1) 90 deg. side
- Press {ENT} (HOLD) key to execute measurement.
- {V2: 270.?}will appear on the display (precisely sight target in face 2) 270deg.side
- Press {ENT} (HOLD) key to execute measurement.
- {DIF 0.00.??} new value {AC: 0.00.??}
- Press {ENT} (HOLD) key to store new correction
- Adjustment is complete.
- Nikon DTM-500 series 520/530/550
- Power unit on and have in normal operation mode.
- Press the (MENU) key.
- Press the (#7) key Collimation program.
- {VR: 90.? } will appear on the display (precisely sight target in face 1) 90 deg. side
{HR.00.? }horizontal value {XR 03”} tilt values {YR 01”} example values - Press (ENT) key to execute measurement.
- {VL: 270.?}will appear on the display (precisely sight target in face 2) 270deg.side
{HR:180.?}horizontal value {XR 02”} tilt values {YR 01”} example values - Press {OK} (ENT) key to execute measurement.
- Adjustment is complete.
- This Collimation procedure for this instrument will adjust the
- Tilt sensors run center and set a software correction for the
- Double centering of the reticule.