Vertical Zero Datum Program for
Sokkia Instruments
Sokkia Instruments
- Sokkia SET 2/3/4 B II
Setup steps may vary depending on year model. You may have to call in.
- Setup instrument and level precisely turn on power and index.
- Press the (2) Prog. Key
- Press the (2) Prog. Key {Correction}
- Press the (1) Menu Key {Collimation}
- Sight the point on 90deg side {Face 1} {Yes/No}
Press (Ent/Shft)<yes> - Sight the point on 270deg side {Face 2} {Yes/No}
Press (Ent/Shft)<yes> - {Observe end} {Yes/No (repeat)} Press (Ent/Shft)<yes>
- {New value set} { Yes/No (exit) } Press (Ent/Shft)<yes>
- Adjustment is complete.
- Sokkia Set 3/4/5 A’s and E’s small frame with 5 buttons.
The only adjustment that you can make in the field is the run Center of the compensator. Which the X value applies correction to the vertical circle.
- Setup instrument and level precisely turn on power and index.
- At the main screen: {EDM THEO S-O CONF}
F1 F2 F3 F4 ESC - Press (F4) {CONF} key
- Press (F2) {Tilt correction}
- Press (F1) {Oset} key this will zero the horizontal angle.
- Press (F1) {SET} key at the zero position.
- Rotate the instrument until the {HAR = 180.00.00}
- Press (F1) {SET} key at the 180 position
- Adjustment is complete.
- Sokkia SET 100 series Set 2100/3100/4100 & 110’s
- Setup instrument and level precisely turn on power and index.
- Horizontally zero set gun while sighted at the prism
- (ESC) key unit you reach the main screen.
- At the main screen: {Meas MEM CNFG}
F1 F3 F4 - Press (F4) {CNFG} to enter config menu.
- Press (#2) {Instr. Const.}
- Press (#1) {Tilt: X### Y###} always perform tilt adjustment first.
- {Take F1} on display Press (F4) {Yes}
- Rotate the instrument horizontally to 180 position.
- {Take F2} on display Press (F4) {Yes}
- {New values X=### and Y= ###}
- Press (F4) {yes} to store corrected values.
- Instrument should return to {Inst. Const.} screen.
- Press (#2) {collimation} key
- Sight the point on 90deg side {Take F1} {Yes/No}
Press (F4){yes} - Sight the point on 270deg side {Take F2} {Yes/No}
Press (F4) {yes} - {Collimation} {EL values & V off. values } Press (F4) {yes}
- Adjustment is complete.
- SET 500/600 series (The Green machines)
- Setup instrument and level precisely turn on power and index.
- Horizontally zero set gun while sighted at the prism
- ESC) key unit you reach the main screen.
- At the main screen: {Meas MEM CNFG}
F1 F3 F4 - Press (F4) {CNFG} to enter config menu.
- Press (Arrow key) to highlight {Instr. Const.} Press (← enter) key
- Press ( ← ) if {Tilt: X#### Y####} is highlighted . always perform tilt adjustment first.
- {Take F1} on display Press (F3) {OK}
- Rotate the instrument horizontally to 180.00.00 position.
- {Take F2} on display Press (F3) {OK} {New values X&Y}
- Press (F4) {yes} to store corrected values.
- Instrument should return to {Inst. Const.} screen.
- Press (arrow key) to highlight {collimation} Press (← ) key
- Index telescope {Oset}
- Sight the point on 90deg side {Take F1}
- Press (F3){OK}
- Sight the point on 270deg side {Take F2}
- Press (F3) {OK}
- {Collimation} NEW {EL values & V off. values }
- Press (F4) {YES} to store the new values.
- Adjustment is complete.
- If Adjustment is out of range you will need to send in for service
- The Vertical tolerance is 30”.